The Step Towards Complete Automation in Sample Preparation
The HPLC Direct Injection module enables the complete automation from raw extract to chromatogram without any manual working step and with excellent results at the same time. Without the need for supervision, samples can be processed during the night or even at the weekend and the finished chromatograms are available the very next morning.
Complete automation of an analytical routine processThe sample preparation is done via the SPE module (Solid Phase Extraction), via the SPE module with online EVAporation or via the GPC module (Gel Permeation Chromatography) and the EVAporation module. The HPLC Direct Injection module can also be used after a solvent exchange with the EVAporation module. |
FREESTYLE HPLC Direct Injection module |
Afterwards the finished samples are injected into the HPLC by sample loop overfilling. The injection volume can be parametrised. The filling of further aliquots of the sample is also possible.
No additional HPLC autosampler is required when using the HPLC Direct Injection module; this function is taking over by the FREESTYLE system that is compatible with any HPLC instrument.
Delivery
Cooled positions for 1 mL or 4 mL vials |
The module consists of ten positions for standards cooled by a Peltier element. Optionally 1 mL or 4 mL vials may be used.
The size of the sample loop is variable; a 20 µL sample loop made of stainless steel is included in delivery as a standard |
Results
In the following we show you some recoveries and a chromatogram of an automated standard SPE process via FREESTYLE SPE with HPLC Direct-Injection-Module:
25 mL of the sample (standard) were added onto a 3 mL AflaCLEAN immunoaffinity column. For the automated sample loading and the subsequent washing step with 10 mL of water, a flow rate of 2 mL/min was parameterised in the FREESTYLE software.
Subsequently the column was dried by a short airflow.
Afterwards the toxin was eluted with 2 mL of methanol into 4 mL-vials. Finally the robotic system injected the eluate from the vial directly into the HPLC by sample loop overfilling.
The result is the ready chromatogram!
B1 | B2 | G1 | G2 | |
---|---|---|---|---|
Recovery rates in % (n=5)* | 98 | 95 | 102 | 85 |
Relative standard deviation in % | 5 | 3 | 5 | 8 |
*Aflatoxin-standard 10 ng / 25 mL
Aflatoxin-standard (10 ng / 25 mL) processed via FREESTYLE SPE with HPLC Direct-Injection Module and a standard SPE-process